Performance evaluation of magnetic bead nucleic acid extraction in real-time fluorescence quantitative PCR detection system for plasma EBV-DNA

Yuhong ZHENG,Fengjin ZHONG,Yansong CHEN,Yingying LIN,Zhenzhou XIAO,Jianji PAN,Shaojun LIN,Qiaojuan GUO,Yan CHEN
2017-01-01
Abstract:Objective To evaluate the application performance of magnetic bead nucleic acid extraction method in realtime fluorescence quantitative PCR detection for plasma Epstein-Barr virus (EBV) DNA.Methods Automated magnetic bead method was used to extract nucleic acid,and combined real-time fluorescence quantitative PCR assays for quantitative EBV-DNA.The detection performance of the method was evaluated byusing the third-party reference materials.In addition,100 patients with NPC were selected,and the plasma EBV-DNA was quantitatively detected by magnetic beads method and boiling method.The correlation between the two methods was analyzed.Results In the automated magnetic bead method-based real-time fluorescence quantitative PCR detection system for EBV-DNA,the repeatability precision of the low,medium and high were all <3/5 TEa (TEa =target ± 0.4 lg),and the intermediate precision were all < 4/5 Tea.In the range of 5.4×101-5.4×105 U/mL,there had good linear relationship.The minimum detection limit was about 3.334×101 U/mL.And the concentration of Hb interference in plasma was lower than 8.667 g/L,the results of magnetic bead test were not significantly affected.The detection results between magnetic bead method and boiling method were linearly related,and the qualitative coincidence rate was 83%.Conclusion The detection of plasma EBV-DNA by automated magnetic bead nucleic acid extraction method can achieve good detection performance.