Active-passive joint measurement and evaluation method for precision of noise test system

Yi-fan GU,Li-ping WANG,Xu DING,Zhi-yu WANG,Jiong-jiong MO,Fa-xin YU
DOI: https://doi.org/10.3785/j.issn.1008-973X.2020.03.018
2020-01-01
Abstract:The gain of the device under test was increased by setting an active gain device behind a passive device, which suppressed the influence of system non-zero error on the noise test results. Combining the twice noise test results of a single active device and a cascaded active-passive device, the characterization parameter of the measurement and evaluation precision of the system was converted from the noise figure into the insertion loss of the cascaded passive device by employing the noise figure direct-subtraction method. Also, the accuracy of the noise test system was quantitatively evaluated by using the insertion loss test value based on vector network analyzer with higher insertion loss measurement accuracy as a reference. The comparison verification test with the traditional passive measurement evaluation method show that, by the proposed method, the evaluation precision of noise test system based on the Y-factor method was –0.5~0.5 dB in the frequency band of 2~40 GHz, while the evaluation precision of noise test system based on the vector cold source method was –0.3~0.3 dB; the range of evaluation precision fluctuation is less than two times the test uncertainty of the system under test, which is much lower than the fluctuation range of being more than six times the uncertainty by the traditional passive measurement evaluation method.