Simultaneous Determination of K+,Ca2+,Na+ and Cl- in Radish by Ion Selective Electrode Combined with Flow-Injection Asynchronous Injection

A-rong WANG,Yong-sheng LI
DOI: https://doi.org/10.11973/lhjy-hx201607012
2016-01-01
Abstract:A rapid and accurate method for simultaneous determination of K+ ,Ca2+ ,Na+ and Cl- in radish was established by using ion selective electrode coupled with flow-injection asynchronous injection.The optimum conditions were as follows:the flow rate was 2.43 mL·min-1 ;the total ion strength adjustment buffer consists of 80 mmol·L-1 of Tris-H3 BO3 (pH 6.00),0.05 mmol·L-1 of Na+ ,0.5 mmol·L-1 of K+ ,0.01 mmol·L-1 of Ca2+ and 0.57 mmol·L-1 of Cl- ;the reaction coil length was 50 cm;the sampling volume was 220 μL;the pH of the mixed standard solution was 6.00.The linearity ranges of K+ ,Ca2+ ,Na+ and Cl- were 10-70,0.5-7.0, 10-36,21 -119 mmol·L-1 ,respectively,with detection limits (3s/k)of 5.0,0.2,5.0,10 mmol·L-1 , respectively.Recovery rates obtained by standard addition method were in the range of 92.8%-110% and RSD (n=11)was less than 2.0%.
What problem does this paper attempt to address?