Optimum b-Value Acquisition Schemes of Diffusion Kurtosis Imaging Based on Brain Microstructural Information Detection

Miao Sha,Yuanyuan Chen,Hongyan Ni,Weiwei Wang,Xin Zhao
DOI: https://doi.org/10.13494/j.npe.20150043
2015-01-01
Nanotechnology and Precision Engineering
Abstract:Diffusion kurtosis imaging ( DKI ) , introducing a higher-order kurtosis tensor to quantify the non-Gaussian property of water diffusion as well as multiple diffusion weighted factor b-values to model fit-ting, is a powerful probe for assessment of the microstructure of biological tissues, enjoying a unique ad-vantage in neuroscience and clinic diagnosis.However, there is still a lack of comparative analysis on different features of microstructure of brain using different b-value combinations in DKI data acquisition and thoughts of optimizing b-value combinations of DKI data acquisition according to different clinical needs.In this paper, DKI image contrast of cerebrospinal fluid, gray matter and white matter and voxel-based analysis of the whole brain were used to compare the influence of DKI parameter under 5 b-values scheme and 2 b-values scheme and the differences of corpus callosum local tissue microstructure charac-teristic expression.Results show that the combination of low b-value (<1 000 s/mm2 ) and high b-value (>2 000 s/mm2 ) DKI scheme has great resolution ability for tissues, while more b-values collected at&nbsp;the same time increase the acquisition time and the model fitting error as well;2 b-values scheme ( 1 000 s/mm2 , 2 000 s/mm2 ) will be more suitable for general DKI clinical diagnosis.
What problem does this paper attempt to address?