Research on test methods of liquid's micro/nano flowrate

Cong Zhang,Sandong Yang,Nai-jie Li,Tao Tang,Yuanshe Sun,Weibing Zhang,Tong Li
DOI: https://doi.org/10.3969/j.issn.1001-232x.2014.05.002
2014-01-01
Abstract:Testing for micro-nano flow rate is one of important works in the μ-HPLC’s development . A new micro-flow rate meter testing method was proposed and evaluated .Its principle was to acquire liq-uid’s instant flow rate by monitoring the temperature change of the fluid .The flow rate was splited by split ratio to achieve the required micro-nano flow rate .The micro-flow rate meter was calibrated by tradi-tional testing method .The results indicated that the micro-flow rate meter testing method possessed better accuracy and repeatability .Besides ,the new method could measure accurate instant flow rate accurately , and the range was μL/min to nL /min .This new method provided an effective way to design and improve micro-nano pumps .