Interaction Between Gold Nanoparticle and Ferrocene in Glucose Biosensors

WU Bao-yan,WANG Yan-yan,YANG Yu,SONG Zhao,YIN Feng,CHEN Qiang
DOI: https://doi.org/10.3969/j.issn.1000-9787.2005.10.004
2005-01-01
Abstract:With the decrease of the applied potential,current response of the glucose biosensor based on gold nanoparticle decreases largely,so ferrocene is used to construct it in order to reduce the amount of current response decreased.Glucose biosensors based on gold nanoparticle with ferrocene as mediator has been formed with glucose oxidase(GOD) immobilized by glutaradehyde in the matrix of glutin on the screen-printed gold electrode,and the interaction between gold nanoparticle and ferrocene is described.The experiments show that ferrocene reduces drastically the amount of current response decreased which is caused by the applied potential,gold nanoparticle reduces the redox potential of ferrocene,and gold nanoparticle and ferrocene can improve the current response of glucose biosensor as a synergetic effect.
What problem does this paper attempt to address?