Research on Flow-Acoustic Coupling of Transducer Disturbing Effect On Ultrasonic Flowmeter

Wen ZHAO,He-ming HU,Shang-chun FAN
DOI: https://doi.org/10.3969/j.issn.1000-8829.2018.05.021
2018-01-01
Abstract:The disturbance flow caused by transducers is an important source of systematic deviation of ultrasonic flowmeters.In order to explore mechanism of the size and configuration of the transducer disturbing effect and calculate laboratory flow calibration factor,the flow-acoustic coupling simulation of a small-diameter flowmeter is introduced.The systematic deviation was obtained by quantitative analysis of velocity distribution along acoustic path,acoustic pressure signal on transducer surface,and the velocity measured by flowmeter.In addition,the systematic deviation under longer acoustic path length was further deduced using a piecewise weighted average method.