Determination of acetone content in PMMA/acetone coating solution by Raman peak contrast method

Nan-nan ZHANG,Zhen-rong ZHENG,Chen-chen HAN,Xiao-ming ZHAO
DOI: https://doi.org/10.16606/j.cnki.issn 0253-4320.2018.01.049
2018-01-01
Abstract:The control of solvent concentration in polymer coating membranes is an important index of film forming quality.The confocal scanning imaging laser Raman spectrometer has the advantages of high resolution,no direct contact with samples,convenient operation and fast speed in the measurement of liquid samples,and is suitable for testing the solvent concentration in polymer coating solution during drying and curing processes.But during the detection,the intensity of Raman scattering light can be easily affected by the background noise and refractive index of solvent in the sample to be measured,the stability of detector and the structure of scattering pool.In order to weaken or offset these influences,the Raman peak strength contrast method is employed to offset the nonlinear relation to a certain extent.In this paper,the content of acetone in polymethyl methacrylate (PMMA)-acetone coating solution is detected by Raman spectroscopy.The results show that Raman contrast method can effectively determine the content of acetone in PMMA/acetone coating solution.When the mass fraction of acetone increases from 52% to 100%,the mass fraction of acetone (Wa) will have a linear relationship with the relative peak-to-peak strength ratio (A1).The linear regression equation is obtained as follows:A1 =-11.371 92+0.229 71· Wa,the linear correlation coefficient outstrips 96%,representing a good fitness.