Determination of Trace Sulfur Compounds in Environmental Air by GC/FPD Coupled the Air Preconcentrations

Xiu-e SHEN,Miao CHANG,Bao-xian LIU,Hong-shuai ZHAO,Xiao-ju WANG
DOI: https://doi.org/10.19316/j.issn.1002-6002.2015.06.018
2015-01-01
Abstract:GC/FPD coupled the Air Preconcentrations was applied to the determination of 7 kinds of trace sulfur compounds in environment air. It was found that the system blank of the Air Preconcentration-GC/FPD was clear, and no residuum after the analysis of sulfur compounds. Results showed that the correlation coefficient of 7 sulfur compounds were all above 0. 995, it has high accuracy and precision. In addition, the recovery of sulfur compounds were in the range of 98. 6%-105. 3%, 99. 1%-103. 7%,85. 7%-99. 7%. The relative standard deviations of the methods lower than 6. 9% for 6 compounds, but for hydrogen sulfide, the relative standard deviation was 10. 9%. Otherwise, when inject volume was 400 mL,the limit detection of 7 sulfur compounds was 0. 10-1. 1 ng/m3 . So the measurement system was sensitive for sulfur compounds and very suitable for determination of the trace sulfur compounds in environment air.
What problem does this paper attempt to address?