Application of Adaptive Pressure-Driven Microfluidic Chip in Thyroid Function Measurement

Xingshang Xu,Nongyue He
DOI: https://doi.org/10.1016/j.cclet.2021.01.008
IF: 9.1
2021-01-01
Chinese Chemical Letters
Abstract:The improvement in accuracy of in vitro diagnosis has always been the focus of early screening of thyroid dysfunction.We constructed a microfluidic chip based on a polystyrene polymer substrate.Total triiodothyronine(TT3),total thyroxine(TT4),free triiodothyronine(FT3),free thyroxine(FT4),and thyrotropin(TSH) in human whole blood samples were analysed by fluorescence immunoassay to evaluate thyroid function.The results indicate that the microfluidic chip shows a good linear relationship in the detection of TT3,TT4,FT3,FT4,and TS H standards,and the correlation coefficient(r) is not less than0.9900.In addition,the chip also has strong anti-interference(RSD%≤5%) and good repeatability(CV≤8%),and its inter-batch differences are small(CV ≤15%).The results of practical application in clinical thyroid function mea surement indicated its high accuracy(r≥0.9900).It provides a new method for the determination of thyroid function and lays a foundation for subsequent clinical application.
What problem does this paper attempt to address?