On the Quality Control for the Determination of Ultratrace-Level <SUP>236</SUP>U and <SUP>233</SUP>U in Environmental Samples by Accelerator Mass Spectrometry

Mu Lin,Jixin Qiao,Xiaolin Hou,Robin Golser,Karin Hain,Peter Steier
DOI: https://doi.org/10.1021/acs.analchem.0c03623
IF: 7.4
2021-01-01
Analytical Chemistry
Abstract:Our recent attempt to determine ultratrace-level U-236 and U-233 in small-volume seawater samples was challenged by high and unstable procedure blanks in our environmental radioactivity laboratory, which used to be a spent fuel research facility. Through intercomparison experiments with different laboratories and background checks on the chemical reagents and laboratory dust, the resuspended U-bearing dust was identified as the dominating source of the U-236 and U-233 contamination. With the implementation of background control (especially dust control) measures, the procedure blanks and detection limits of U-236 and U-233 for the radiochemical separation procedure have been significantly improved by three orders of magnitudes. With well-controlled blanks, the analytical precision for U-236 and U-233 predominantly relies on the AMS counting statistics. Background check and dust control are strongly recommended before the analyses of environmental-level long-lived radionuclides (such as U-236 and U-233) that are conducted in the former or active nuclear facilities, even if clearance of radioactivity relevant for radioprotection was achieved.
What problem does this paper attempt to address?