Abstract:Objective: To assess the technical feasibility and effectiveness of fluoroscopically-guided self-expanding metallic tracheobronchial Y-shaped stents implantation under local anaesthesia by introducing two stiff guidewire at one time for shortening the procedure time. Methods: A total of 16 self-expanding metallic tracheobronchial Y-shaped stents implantation procedures were retrospectively enrolled in this study. In 6 procedures the two stiff guidewire were routinely introduced respectively. In another 13 procedures the two stiff guidewire were introduced at one time. The technical success rate, average procedure time was documented. Results: 19 self-expanding metallic tracheobronchial Y stents were all successfully implanted in treatment of malignant tracheal carina stenosis. The average procedure time of routine group was 15.7 mins, while the technique modification group was 11.0 mins. The difference between the two was significant (P = 0.001,2). Conclusion: The two stiff guidewire introduced at one time can shorten the procedure time significantly in self-expanding metallic tracheobronchial Y stents implantation.