A Novel Wavy Non-Uniform Ligament Chiral Stent with J-shaped Stress–strain Behavior to Mimic the Native Trachea
Liu, Jiapeng,Yao, Xinhua,Wang, Zhenwei,Ye, Jian,Luan, Congcong,Fu, Jianzhong,He, Yong
DOI: https://doi.org/10.1007/s42242-021-00159-y
2021-01-01
Bio-Design and Manufacturing
Abstract:Tracheal stents are an important form of treatment for benign or malignant central airway obstruction. However, the mechanical behavior of current tracheal stents is significantly different from that of the native trachea, which leads to a variety of serious complications. In this study, inspired by the structure of the native trachea, a wavy non-uniform ligament chiral tracheal stent is proposed, in which J-shaped stress–strain behavior and negative Poisson’s ratio response are achieved by replacing the tangential ligament of tetrachiral and anti-tetrachiral hybrid structure with a wavy non-uniform ligament. Through the combination of theoretical analysis, finite element analysis and experimental tests, a wide range of desired J-shaped stress–strain curves are explored to mimic the native porcine trachea by tailoring the stent geometry. Besides, the negative Poisson’s ratio and auxetic diameter curves versus axial strain of the stent are also studied in detail, thus contributing to the enhancement of cross-section ventilation and reducing the migration of the stent. This novel tracheal stent with a unique microstructure shows a potential to perfectly match the physiological activities of the native trachea and thereby reduce potential complications.