Pathlength selection of determing tire chemical oxygen demand (COD) in wastewater by using near-infrared transmission spectra

He Jin-Cheng,Yang Xiang-Long,Wang Li-Ren
2007-01-01
JOURNAL OF INFRARED AND MILLIMETER WAVES
Abstract:The chemical oxygen demand ( COD) in wastewater can be determined by near-infrared spectra (NIRs), however it is strongly influenced by the water absorbance. It is then very important to choose proper pathlength for improving measurement sensitivity and thus minimize the measurement error. In this study, we presented a series of experiments to determine the optimum pathlength for COD determination using a Fourier transform near infrared instrument with transmission measurement. Sixty wastewater samples for this study were collected from hoggery. Partial least squares ( PLS) models were developed based on the spectral data which had been collected by using the cell pathlength of 1 mm, 2mm, 5mm, and 10mm. The performances of the differents models were compared. The results show that the pathlength of 10mm developes the optimal performance in the spectral range 800-1350nm and that of 1 mm or 2 mm of pathlength can be obtained in the spectral range of 1350 similar to 2500nm. In addition, our results suggest that two pathlength combination can improve the reliability and accuracy of the model in the region of 800 similar to 2500nm. This method of two pathlength combination may also be applied to the quantitative analysis of complex components in aqueous solutions.
What problem does this paper attempt to address?