COD Detection in Wastewater Treatment Based on Three-dimensional Fluorescence Spectrometry

DU Shu-xin,WU Yuan-qing,WANG Jian
DOI: https://doi.org/10.3969/j.issn.1000-3932.2013.08.006
2013-01-01
Abstract:The three-dimensional fluorescence spectrometry was employed to analyze COD detection in the primary, secondary and final sedimentation tanks of a wastewater treatment plant;and the performance of calibration models respectively established with principal component regression(PCA),the partial least squares(PLS),the parallel factor analysis(PARAFAC) and the multi-way partial least squares(N-PLS) were compared to show that the conventional un-folded PLS method outperforms three-dimensional modeling methods like PARAFAC and N-PLS in COD detection.
What problem does this paper attempt to address?