The effect of different reconstruction algorithms on the image quality of SPECT Hoffmann phantom

王嘉兴,刘丛进,朱汇庆,张锦明,刘兴党
DOI: https://doi.org/10.3760/cma.j.issn.0253-9780.2010.02.017
2010-01-01
Abstract:Objective To evaluate the image quality of SPECT Hoffmann phantom using different reconstruction algorithms.Methods Data sampling of line-source and Hoffmann phantom were acquired with parallel-hole low energy and high resolution collimator by a dual-head SPECT system.The images were reconstructed by both filtered backprojection (FBP) and ordered-subsets expectation maximization (OSEM) algorithms.The spatial resolution of line resources was evaluated by the full width at half maximum (FWHM) value.The image quality of Hoffmann phantom was visually accessed by 3 doctors.The reconstruc-tion time was recorded and region of interest (ROI) was indentified on the reconstructed Hoffmann phantom images.Independent samples t-test was used to compare the ratios of counts/pixel of target ROI to cerebellum ROI by different reconstruction algorithms using software SPSS 15.0.Results The FWHM values for FBP and OSEM were 18.77 mm and 12.62 mm, respectively.The visual assessment suggested that OSEM could provide better overall image quality, better edge delineation of candate, putamen and striatum than FBP.The reconstruction time of FBP and OSEM was 80 s and 100 s respectively, which was acceptable in routine clini-cal procedure.The ratios of counts/pixel of target ROI to cerebellum ROI generated by FBP and OSEM showed no significant difference (t=-0.332, P=0.750).Conclusion OSEM algorithm could provide bet-ter spatial resolution and image quality than FBP with an acceptable increase in reconstruction time.