Dosimetry research of different radiation therapy technology for patients whit medulloblastoma

李夏东,吴稚冰,马胜林,章霓,邓清华,赵春雷,唐荣军,赖建军
DOI: https://doi.org/10.3760/cma.j.issn.1004-4221.2013.01.022
2013-01-01
Abstract:Objective To research on the dosimetry of medulloblastoma three-dimensional radiotherapy technology using linear accelerator and comparing it with helical tomotherapy.Methods From January 2011 to April 2011,12 patients with medulloblastoma were scanned in CT-simulator in separate image sequences,of which all sequences were coupled to complete sequences.The 12 patients with medulloblastoma were used to generate 7 beams three-dimensional conformal radiotherapy (3DCRT),9 beams,13 beams intensity-modulated radiotherapy (IMRT) plans.36 Gy were prescribed to the planning target volume (PTV) of brain and spinal and the posterior fosse tumor bed added amount to 54 Gy.The dose parameters of PTV and organs at risk used linear accelerator were compared with helical tomotherapy.Results 3DCRT and IMRT require three and two treatment centers,it will makes the cranial nest in low dose region and dose vacuum in beam cohesion area when using 3DCRT technology.The dose homogeneity index of PTV using 9 beams IMRT was better than 3DCRT and 13 beams IMRT (0.93,0.82,0.89,F =6.17,P =0.02).The dose conformity index of PTV using 9 beams IMRT was better than 3DCRT and 13 beams IMRT (0.97,0.88,0.95,F =5.23,P =0.01).The dose distribution was similar with 9 beams IMRT and tomotherapy.Conclusion It is recommend to use image coupling method using conventional linear accelerator with 9 beams IMRT plans which will led to a similar dose distribution compared with helicon tomotherapy.