Comparison of Success Rate of the Self-Drilling Microscrews Placed by Self-Drilling or Pre-Drilling Techniques

SHU Guang,TANG Zhi-hui
DOI: https://doi.org/10.3760/cma.j.issn.1674-5760.2010.01.007
2010-01-01
Abstract:Objective To compare the success rates of the self-drilling miniscrews placed by self-drilling or pre-drilling techniques. Methods 272 miniscrews placed by self-drilling or pre-drilling in 117 orthodontic patients were included. Chi-square tests and multiple logistic regression analyses were performed to compare the success rates and possible influencing factors. Results The overall success rate was 86%. The success rates of miniscrews placed by self-drilling and pre-drilling techniques were 85. 3% and 86. 8%, respectively. The difference did not reach statistical significance.Multiple logistic regression analyses revealed that the factor influencing the success rate of pre-drilling technique was the combination of implant position and skeletal pattern, while the factors influencing the success rate of self-drilling technique were both the implant position and the combination of age and skeletal pattern of the patient. Conclusions Self-drilling technique is preferred in most situations,especially for maxillary miniscrew placements for high angle patients, however, for adult low angle patients or for mandibular miniscrew placements, pre-drilling technique is preferred.
What problem does this paper attempt to address?