Biometric measurement by IOL-Master in cataract patients with high myopia

Juan ZHANG,Jing WANG,Han ZHANG,Jing LIU,Guoling CHEN,Xue GAO,Li-xia YANG
DOI: https://doi.org/10.3760/cma.j.issn.1006-4443.2010.08.021
2010-01-01
Abstract:Objective To compare the measurement of axial length (AL) by IOL-Master and contact ultrasonic (US) axial scan (A-scan) in high myopia with cataract. Methods The data of 18 patients (28 eyes) who were all high myopia with cataract and accepted the phacoemulsification and foldable IOL implantation in our hospital from September in 2008 to September in 2009 were analyze.The measurement of axial length was performed in 28 eyes by contact A-scan and IOL-Master, respectively. Keratometric power was measured by IOL-Master. And IOL power calculation was carried out according to SRK/T formula in the basis of IOL-Master. The cases were followed up for one month. Results The axial length was (28.71±1.14)mm by contact A -scan, (29.31±1.27)mm IOL-Master, respectively. There were significant difference between contact A-scan and IOL-Master (P <0.05). Conclusions The IOL-Master is more accurate than the contact A-scan, and IOL-Master is an better choice to measure the axial length in high myopia with cataract.