Comparison of anterior segment biometry and intraocular lens power measured by Lenstar LS900 and IOLMaster

YE Xiang-Yu,JI Jian-Li,ZHANG Guang-Bin,WANG Le,MAO Yi-Hui,HE Xue-Hong
2011-01-01
Abstract:Objective To compare the difference of anterior segment biometry and intraocular lens(IOL) power measured by a new noncontact optical biometer Lenstar LS900(Haag-Streit) using optical low-coherence reflectometry and IOLMaster.Methods Biometry measurements in 60 eyes of 30 patients scheduled for cataract surgery by the same examiner with Lenstar LS900 and IOLMaster,the axial length(AL),anterior chamber depth(ACD) and keratometric(K) were compared,IOL power for AcrySof SA60AT(Alcon)measure by two devices using SRK/T formulas was calculated,and the difference of targeted diopter under same IOL power was compared.Results There was no statistical difference in AL,ACD,average K and steep K measured by Lenstar LS900 and IOLMaster(all P>0.05),and they also showed the excellent relation(r=0.999,0.955,0.936,0.845) except for the steep axis(r=0.579).The targeted diopter calculated with Lenstar LS900 and IOLMaster were(-0.02±0.17) D and(-0.07±0.23)D,respectively,there was no statistical difference(P>0.05).Bland-Altman plots showed excellent agreement in IOL power measurements as determined by SRK/T formulas by Lenstar LS 900 and IOLMaster.Conclusion The anterior segment biometry and IOL power measured by Lenstar LS900 and IOLMaster have better agreement.Lenstar LS900 is an accurate and fast instrument,which can provide the additional information for any cataract or refractive surgeon.