A Comparative Study Between the Humphery Matrix N-30-F and FDT N-30 Tests

HE Min,WU Ling-ling,XIAO Zhen,HUANG Ping
DOI: https://doi.org/10.3760/cma.j.issn.1006-4443.2011.02.0
2011-01-01
Abstract:Objective To determine the comparability of the N-30-F test with the new Humphrey Matrix instrument to the traditional frequency-doubling technology(FDT)N-30 test.Methods A sample of 32 eyes from 32 patients(3 normal subjects,6 with ocular hypertension,10 with primary open-angle glaucoma (POAG),13 with suspect glaucoma)underwent visual field testing with standard automated perimetry,and with FDT,using the FDT N-30 program,and the Matrix N-30-F program.The plots,mean deviation(MD),pattern standard deviation(PSD),defect locations on the Total Deviation Probability(TDP)and test time were compared.Results No statistically significant differences were found between corresponding parameters of the two tests.High coefficients and the agreement were associated with the MD values,the PSD values and the total number of abnormal test locations on the TDP plot of the two tests.The test time for the Matrix test was significantly shorter.Conclusions The FDT N-30 and Matrix N-30-F tests are comparable in screening for early diagnosis of glaucoma,and the test time for the Matrix test is significantly shorter.