Influence Factor on Virtual Calibration Result of Lung Counter

Zhen Wu,Rui Qiu,YiZheng Chen,ChunYan Li,JunLi Li
DOI: https://doi.org/10.7538/yzk.2017.51.03.0529
2017-01-01
Abstract:In the in vivo measurement for internal contamination , the detector count from the radionuclides deposited in organs other than the lung is neglected for lung counter .Besides ,the monitored people have morphological deviation with the reference phantom in calibration .The organ-specific detection efficiency for 65Zn and 60Co nuclides was simulated with Monte Carlo code by using the Chinese adult male phantom library , and the total detection efficiency was calculated .The effect of organ-specific detection efficiency and total detection efficiency on virtual calibration was studied .The results show that the contribution of nuclides like 65 Zn in the organs other than the lung can not be neglected in the virtual calibration .T he total detection efficiency of the lung counter for 65 Zn and 60 Co nuclides decreases with the increase of weight at the same height ,and the decreased proportion for 65 Zn varies with the time after intaking .For each phantom , the total detection efficiency decreases with the increase of time .The results prove that the radionuclide biokinetic distribution and morphological variation have a great influ-ence on the total detection efficiency of the lung counter ,w hich varies with nuclides .
What problem does this paper attempt to address?