Research for PET imaging divided by time-based and amplitude-based respiratory gating

苏畅,王瞳,邢海群,霍力,李方,张辉
DOI: https://doi.org/10.3969/J.ISSN.1672-8270.2017.03.001
2017-01-01
Abstract:Objective:To explore the effectiveness of positron emission tomography (PET) respiratory gating based on time and amplitude to improve the reconstructed image artifacts.Methods:Two respiratory gating methods were compared based on 8 clinical patients data, equal intervals of time division method time-based gating and equal data division method variable amplitude-based gating, the maximum displacement, maximum standardized uptake value (SUVmax) and gross tumor volume were used to quantitatively evaluate these two methods. Independent-t test was used to test whether the difference between two methods were significant for SUVmax and gross tumor volume.Results: For patient data involving heart, amplitude-based method detected 24.47% more motion signal which related to maximum displacement than time-based method. For patient data involving lung lesions, SUVmax derived from amplitude-based method could be recovered to a higher value (9.92% more than time-based method), and 40% (4/10) of the patient data results were statistically significant (t=-3.805,t=-9.099,t=-2.817,t=-2.315,P<0.05). Gross tumor volume calculated from amplitude-based method could be 29.13% less than time-based method, and the differences were statistical significant(t=2.262,t=10.382,t=2.922,t=2.762,t=2.858,P<0.05), besides the percentage of patient data was 50% (5/10).Conclusion:Amplitude-based method can extract more motion information, at the same time, signal to noise ratio values of different phases' images are similar to each other.