The Study on Acquisition Mode and Reconstruction Parameters of Brain FDG PET Images

左传涛,刘永昌,管一晖,赵军,林祥通
DOI: https://doi.org/10.3760/cma.j.issn.2095-2848.2001.06.019
2001-01-01
Abstract:Objective To evaluate the effect of acquisition mode on the brain PET images. Methods After changing conditions and parameters,we got brain PET images of different acquisition modes,different emission counts,different transmission times; and compared with the reference images the impacts of different acquisition modes,different acquisition conditions were assessed. Results Compared with 2D mode, much higher background and noise were observed on the reconstruction images of 3D mode, and the bottoms of the brain structure were not well displayed. But the middle part of brain structure displayed well in 2D and 3D mode without difference; the gray/white radioactivity ratios were 2.108±0.183 and 2.286±0.232 under 2D and 3D mode,respectively. The gray/white radioactivity ratios with different emission counts were 2.108±0.183, 2.125±0.158, 2.161±0.176, respectively, there was no evident difference among them. With transmission counts increasing, the segmented image outline of Hoffman phantom and brain structure became clear and integral. Conclusions Different acquisition modes, different emission counts and different transmission times are of certain impacts on brain FDG PET images, and it should be paid more attention in clinical practice.
What problem does this paper attempt to address?