Improving Detection Sensitivity of a Low Background BEGe Spectrometer by Pulse Shape Discrimination Using Rise-Time Ratio

Yu-Hao Mi,Zhi Zeng,Hao Ma,Jian-Ping Cheng,Li,Hui Zhang
DOI: https://doi.org/10.1007/s10967-020-07191-9
2020-01-01
Journal of Radioanalytical and Nuclear Chemistry
Abstract:To improve detection sensitivity for interesting radionuclides in measurements of environmental samples by Broad Energy Germanium detectors, the Pulse Shape Discrimination method using rise-time ratio is introduced. A certified soil sample and a typical squid-ash sample were measured by a well-shielded BEGe spectrometer. Minimum Detection Activities for the 722.9 keV and 949.8 keV peaks from the squid-ash sample were decreased by about 26% and 35% respectively. The method performs well especially when the main background comes from gamma radionuclides in samples.