Mismatch in size of occluding devices for atrial septal defect ( ASD ) impairs endothelialization of the occluder surface in a canine ASD model

Ge-Sheng Cheng,Yu-Shun Zhang,Ting-Ting Zhang,Chen Wan,Xu-Mei He,Ya-Juan Du
2016-01-01
Abstract:Endothelialization of the atrial septal defect (ASD) occluding device is of major clinical relevance. We aim to identify the effect of occluding devices of different sizes on endothelialization of the occluder surface in a canine ASD model. A 6 mm-diameter circular defect was generated in the center of the oval fossa of 12 dogs. The animals were then randomly assigned to receive an 8 mm-diameter occluder (group A), a 12 mm-diameter occluder (group B) or a 16 mm-diameter occluder (group C). The shape of the occluders and endothelialization of the occluder surface was assessed by different methods. Transthoracic echocardiography (TTE) showed that no occluder dropped off or changed position. Fourteen months after implantation, the occluder surface was completely covered with endothelial tissues in group A, while in group C the occluder surface was covered with islands of endothelial tissues. Immunoblotting assays showed the highest endothelial expression of eNOS in group A (1.21±0.25) followed by group B (0.73±0.18). The nuclei were clearly seen and evenly distributed and tight junctions were observed between endothelial cells in group A on transmission electron microscopy. The occluder surface in group C showed only partial coverage by endothelial tissues and the center of the occluder only showed skeleton of collagen fibers. Transmission electron microscopy showed lack of tight junction between endothelial cells and gaps were present. Oversized occluders show impaired endothelialization of the occluder surface.
What problem does this paper attempt to address?