Tripodal Squaramide Derivative As a Neutral Chloride Ionophore for Whole Blood and Sweat Chloride Measurement

Lingling Fan,Tingting Xu,Junjun Feng,Zihan Ji,Le Li,Xinhao Shi,Chunxiu Tian,Yu Qin
DOI: https://doi.org/10.1002/elan.201900693
IF: 3.077
2019-01-01
Electroanalysis
Abstract:A tripodal squaramide derivative was synthesized and characterized as a new neural ionophore for chloride ion. Squaramide based compounds are known to be better hydrogen bonding receptors than urea derivatives because they donate four hydrogen bonds to anions, while the tripodal structure provides a 3D geometry optimum for spherical anion complexation. The ionophore demonstrated high selectivity over salicylate and heparin that are major interferences for chloride measurement in biological samples with direct potentiometric method. The new ionophore based solid-contact electrodes were fabricated and showed the measuring range of 10(-5) to 1 M chloride ion in the presence of 1 mM of salicylate and heparin with near Nernstian slope of 55 mV/decade. The sensor was successfully applied for the chloride measurement in undiluted heparinized whole blood and artificial sweat samples with our clinical analyzer.
What problem does this paper attempt to address?