Visual and Real-Time Imaging Focusing for Highly Sensitive Laser-Induced Fluorescence Detection at Yoctomole Levels in Nanocapillaries.

Wenmei Zhang,Lei Liu,Qi Zhang,Dongtang Zhang,Qin Hu,Yanan Wang,Xiayan Wang,Qiaosheng Pu,Guangsheng Guo
DOI: https://doi.org/10.1039/c9cc09594b
IF: 4.9
2020-01-01
Chemical Communications
Abstract:A highly sensitive laser-induced fluorescence detection system with real-time imaging focusing instead of the use of fluorescent reagents were developed for the detection of analytes in nanocapillaries. A cylindrical lens was adopted for the shaping of a laser beam to increase its spatial resolution. The limit of detection with a capillary having a radius of 440 nm was 6.76 yoctomoles (or four molecules) for fluorescein, and 84.5 yoctomoles (or 51 molecules) for FAM-labeled Hsa-miR-17. The designed system exhibited high sensitivity and stability for femtoliters of sample and was hence indicated to be suitable for analyses of single cells.