An Auto-Correction Laser Photoacoustic Spectrometer Based on 2f/1f Wavelength Modulation Spectroscopy

Ke Chen,Shuai Liu,Liang Mei,Feng Jin,Bo Zhang,Fengxiang Ma,Yewei Chen,Hong Deng,Min Guo,Qingxu Yu
DOI: https://doi.org/10.1039/c9an01799b
2020-01-01
Abstract:An auto-correction laser photoacoustic (PA) spectrometer based on 2f/1f wavelength modulation spectroscopy (WMS) has been proposed and demonstrated for trace gas detection to eliminate concentration measurement errors due to light power variations. A 1.53 μm distributed feedback (DFB) laser is used as a light source to excite the 2f PA signal that is generated by gas absorption. Meanwhile, a multilayer graphene sheet is employed as a highly efficient photothermal conversion unit for adding a 1f PA background signal, whose amplitude at the absorption center of the traditional PAS system is almost zero. The experimental results show that the gas concentration measured from the 2f/1f WMS signal is almost independent of the laser power. A detection limit of 416 ppb has been achieved for the 1 s measurement interval and could be further improved to 65 ppb with 100 s averaging, according to the Allan deviation analysis.