The Study on the Biodegradation Mechanism of the HA/PLGA Composite Made by High-Pressure for Internal Fixation of Bone Fractures

于家傲,路来金,李玲玲,刘志刚,邹广田
DOI: https://doi.org/10.3760/cma.j.issn.1005-054x.2006.01.021
2006-01-01
Abstract:Objective To investigate the biodegradation mechanism of the HA/PLGA composite rods for internal fixation of bone fractures. Methods The HA/PLGA samples were immersed in the simulated body fluid (SBF) at 37℃. The average molecular weight and the ratio of the losing weight were measured periodically in vitro. In vivo study the samples of HA/PLGA were implanted subcutaneously and in the medullary cavities of News Zealand rabbits . The specimens were examined histologically by light microscope. The rod surfaces were examined using a scanning electron microscope. Results In the initial hydrolysis process of HA/PLGA, HA particles detached from the rod surface and water infiltrated into the rod. The average molecular weight dropped while the weight loss ration increased . In vivo observation showed that the biodegradable material was used by the body. New vessels grew into the samples to facilitate further degradation. Conclusion The HA/PLGA composite is biocompatible. It is also biodegradable and allows new bone formation.