Research on Quick Measuring Instrument of Luminance Uniformity and Color Temperature for Medical Lampbox

周文光,孔悦,姚军
DOI: https://doi.org/10.3969/j.issn.1003-8868.2010.09.024
2010-01-01
Abstract:Objective To develop an imagining-style color analyzer so as to increase measuring accuracy. Methods The tristimulus values of the illuminant were measured with the methods of converting the black-white CCD area array and three colordisc groups, and the height and chromaticity coordinate of the different points were calculated by recording three pictures of the X, Y, Z tricolor of the measured objects. Results The same medical lampbox was measured by a color analyzer and a multi-colored BM-7 respectively. By comparison, the residual of the height was less than 0.2%, with that of the uniformity of luminance less than 0.1% and the difference between the chromaticity coordinate was 0.002. Conclusion The imagining-style color analyzer, which is small and can measure quickly and accurately, is suit to measure color temperature and uniformity of luminance of the surface light source.
What problem does this paper attempt to address?