The Design and Commissioning of the Vacuum-Ultraviolet Circular Dichroism Facility at NSRL

ZHANG Guobin,WU Wenqing,SHI Junyan,ZHOU Hongjun,QI Zeming,FU Yibing,Huo Tonglin
DOI: https://doi.org/10.3321/j.issn:0253-3219.2007.02.005
2007-01-01
Nuclear Techniques
Abstract:This paper describes the design and commissioning of the vacuum-ultraviolet circular dichroism facility at NSRL. At 0.5nm resolution, the normal incident beamline can provide ultraviolet and vacuum-ultraviolet radiation at fluxes higher than 1012ph/s, which is converted into the plane polarized light using a polarizer and subsequently converted into the circularly polarized light using a photoelastic modulator with a switching frequency of 50 kHz. Example spectra of (1s)-(+)-10-camphorsulphonic acid (CSA) and myoglobin are given. Normally, the measurable wavelength range is 178—300 nm. But the short wavelength limit can be shifted to 140nm by using special solvent and short light-pass sample cell.
What problem does this paper attempt to address?