Determination of Dinoprostone in Controlled-Release Dinoprostone (PROPESS) by High Performance Liquid Chromatography

赵轶,国强,虞和永,杨波
DOI: https://doi.org/10.3321/j.issn:1001-2494.2009.08.017
2009-01-01
Abstract:OBJECTIVE: To establish a method for the determination of dinoprostone by HPLC-UV. METHODS: The separation was performed on a Eclipse XDB-C 18 column (4.6 mmx150 mm, 5 μm) with 0.2% acetic acid - aethanol (49:51) as the mobile phase,the detection wavelength was 210 nm and the flow rate was 0.8 mL·min -1. The column temperature was 30°C and the amount of injection was 20 μL. RESULTS: The retention time of dinoprostone was 5.5 min, the linear range of the method was from 24.8 to 248.0 mg·L -1, and the detection limit was 1.24 mg·L -1. The recoveries of dinoprostone were from 99.3%-100.6%, the intraday and interday variations were less than 2.0%. CONCLUSION: The method is simple, accurate, and suitable for routine analysis.
What problem does this paper attempt to address?