Quantitative Analysis of Multi-component with FTIR and Its Application

WANG Hui,ZHAO Xiu-ge,XIAO Wen-de,LU Wen-zhi,YUAN Wei-kang
DOI: https://doi.org/10.3969/j.issn.1006-3080.2001.01.001
2001-01-01
Abstract:The application of FT-IR to simultaneous quantitative analysis of flue gas was studied with the help of Nicolet's Protege 460 combined with a 2m gas cell. Based on CLS method, a CLS measuring method was established which demonstrates the linear relationship between volume fraction and adsorption. Furthermore, the non-linear relationship is calibrated. As a result, the method can simultaneously analyze components in a mixture quantitatively. The equations of the calibration curve for N2O,NO,NO2,SO2 and CO2 at 100°C were thus obtained as follows, N2O: φca=0.707 5φ+1.435 2×10-3φ2 NO: φca=0.719 4φ+3.302 7×10-4φ2 NO2: φca=0.790 7φ+1.986 9×10-4φ2+2.949 7×10-7φ3 SO2: φca=0.991 7φ CO2: φca=0.979 4φ φ is the volume fraction in the mixture, φca is the calibrated volume fraction. Further discussion showed that this is a universal method for infrared absorption gas. The method, with its rapid scan and high sensitivity, can be used in quantitative analysis as well as qualitative analysis of new species when combined with FT-IR technique.
What problem does this paper attempt to address?