Study on the technique for quick determination and whole sampling of VOC in indoor air

TAN He-ping,MA Tian,SUN Deng-feng,XIE Zhen-wei,FANG Zheng,DAN De-zhong
DOI: https://doi.org/10.3969/j.issn.1674-5124.2006.01.001
2006-01-01
Abstract:We sample from indoor air using silonite coated canisters,preconcentrate low level samples utilizing three stages focuser,analyze VOC in indoor air with GC/MS equipment.The standard VOC gas in our study has 39 components.We draw calibration curve with VOC concentrations of around 100,50,25,10,5,2 ppb.The correlation coefficient of the curve is greater than 0.99,the Relative Standard Deviation is below 5%(n=6),and the Medium Relative Standard Deviation is below 5%(n=6).According to the results of sampling and analyzing of indoor air,the relative error of parallel samples is below 10%.The Extended Uncertainty of this method is 10.6%.Using this method we realize the whole sampling and analyzing of VOC,avoiding the error coming from the selective adsorption of the sorbent.
What problem does this paper attempt to address?