A Method for Ouick Determination of Formaldehyde in Indoor Air

He-ping TAN,Zheng FANG,Tian MA,Ling-ping ZENG
2008-01-01
Abstract:A method of determining the formaldehyde in indoor air is studied by using formaldehyde analyzer BH-1. According to our study of typical low, medium and high concentration formaldehyde gases, the relative error is below 1.8%, RSD is below 5.6%(n = 6), and the mean relative standard deviation is below 4.9% (n = 6). Also this method is compared with the national standard method (GB/T 18204.26 Hydroxybenzene reagent spectrophotometer method) at laboratory and on the spot, and the results show no distinct difference between these two methods. The total uncertainty of the BH-1 analyzer analyzing formaldehyde is 10.4%. This instrument method is fast and sensitive, and suitable for determining formaldehyde on the spot, especially when the spots or the sites are numerous. It proves that this fast method can be used with equal efficiency as the hydroxybenzene reagent spectrophotometer method.
What problem does this paper attempt to address?