Abstract:Objective:To evaluate the diagnostic value of spin echo (SE) and GE-short time inversion recovery (GE-STIR) MR sequence in bone injury of knee joint.Methods:The MRI data of 50 cases with knee injury was analyzed retrospectively.All cases were performed with SE sequence (Sagittal T1W,T2W) and Sagittal GE-STIR sequence.The size,signal intensity,margin of bone contusion,MRI features of meniscus,ligaments and fluid were analyzed according to the different MRI Sequence,and also the difference between SE and GE-STIR sequence were compared.Results:Among 50 cases,there were 54 fuci of bone contusion were demonstrated in 45 cases and there were 6 fuci of cartilage injury,31 cases with joint effusion,23 cases with injury of meniscus,5 cases with injury of anterior cruciate ligament,1 case with fibia collateral ligament.43 and 47 fuci were demonstrated on SE T1W,T2W respectively,54 fuci were depicted on GE-STIR.The differences of the size,signal intensity homogeneity,border of bone contusions were statistically higher on GE-STIR than that on SE (P0.05).There were no statistically significant for joint effusion,meniscus injury and ligament injury between SE and GE-STIR imaging (P0.05).Conclusion:GE-STIR Sequence has great diagnostic value in contusion of knee joint,which is superior to SE sequence.