Feedback System for Artificial Electrical Prosthesis Based on In-vivo Impedance Measurement

Hong-liang WANG,Kai-jie WU,Ting LIANG,Xin-yu CHAI,Qiu-shi REN
2009-01-01
Abstract:Objective: To constitute a feedback system for optic nerve prosthesis and choose a method to measure the electrode-optic nerve impedance, which could be used to evaluate the condition of stimulation electrodes. Methods: This paper describes an electrode-optic nerve impedance model firstly, and then adopts three-electrode measurement method to measure the electrode-optic nerve impedance with stimulation current as measurement current. Results: Experimental results show that in the range of allowable error, the method is available in electrode-optic nerve impedance measurement. Conclusion: Electrode-optic nerve impedance could be achieved precisely by micro-stimulator without additional signal synthesizer.
What problem does this paper attempt to address?