Biofeedback Neuromuscular Electrical Stimulation Front-End for Dysphagia Treatment.

Rongtao Hu,Yajie Qin,Yinghong Tian,Zhiliang Hong,Geng Yang,Li-rong Zheng,Jie Jia,Cuiwei Yang,Xiaomei Wu,Yuanyuan Wang
DOI: https://doi.org/10.1109/biocas.2014.6981800
2014-01-01
Abstract:A dedicated front-end for biofeedback neuromuscular electrical stimulation (NESM) system is proposed. For controllable dysphagia treatment, the integrated circuit (IC) provides a stimulator front-end with programmable stimulation parameters (2 mu A-1mA current amplitude, DC-2KHz frequency, and variable duty cycle) and an electromyogram/impedance (EMG/ETI) readout front-end with programmable gain and bandwidth (42-80dB, 0.1Hz-1.2KHz) for biofeedback. Area-efficient, low-power but high precision current controlling is achieved by inducing the sigma-delta modulator technique in the stimulation channel. The measured impedance and EMG signal are used to determine the stimulation parameters, enabling a closed loop optimized treatment. The proposed front-end is fabricated in a 0.18 mu m standard CMOS process technology and dissipates a peak power of 2.3 mW at the supply voltage of 1.8 V. Measurement results on a live person are also provided to validate the system's effectiveness.
What problem does this paper attempt to address?