Visual Prosthesis for Optic Nerve Stimulation

Xiaohong Sui,Liming Li,Xinyu Chai,Kaijie Wu,Chuanqing Zhou,Xiaodong Sun,Xun Xu,Xiaoxin Li,Qiushi Ren
DOI: https://doi.org/10.1007/978-0-387-77261-5_2
2009-01-01
Abstract:The C-Sight visual prosthesis is based on optical nerve stimulation with a penetrating electrode array. A silicon-based microprobe by MEMS process techniques and Pt-Ir microwire arrays by electrochemical etching were fabricated in our project. Noise and impedance analyses were applied to optimize the electrode configuration. A multichannel microcurrent neural electrical stimulator and an implantable CMOS-based micro-camera were developed for neural stimulation and image acquisition, respectively, with a DSP-based system processing the captured image. Electrical evoked potentials (EEPs) from rabbit models were recorded using multichannel stainless-steel screws Mounted oil the primary visual cortex. The mean charge threshold density was 20.99 +/- 5.52 mu C/cm(2) considering the exposed Surface of the stimulating electrode. Current threshold decreased as the pulse duration of the stimulus increased while the corresponding charge threshold increased. The amplitude of PI increased when the Pulse duration increased from 0.4 to 1.0 ms while the latency of PI changed little. Experiments also showed that different distribution maps of EEPs were elicited by different pairs of stimulating electrodes. The stimulating electrode pair along the axis of the optic nerve elicited cortical responses with much lower thresholds than that perpendicular to the axis of the optic nerve. The visual prosthesis with stimulating electrodes penetrating into the optic nerve has been validated in animal experiments.
What problem does this paper attempt to address?