A Water-Cooling Eletrome on Reducing Tissue Thermal Damage

Minggao Li,Xinggang Jiang,Wenshuai Wu,Deyuan Zhang
DOI: https://doi.org/10.1109/icma.2018.8484293
2018-01-01
Abstract:The increase of tissue temperature in monopolar electrome electrosurgery will damage the tissue inevitably. A water-cooling electrome is proposed to minimize tissue thermal damage. The coolant pours to the surgical tissue interface from the water-cooling electrome tip and takes away extra tissue heat. Consequently, the tissue damage decreases. A water-cooling electrome prototype is designed and comparative experiments with ordinary electrome are conducted. The experimental results show that the water-cooling electrome reduces the tissue damage at least 100 um compared with the ordinary electrome at the same output power. When the electrome power reaches 60 W, the average thermal lesion width of water-cooling electrome compared with that of ordinary electrome decreases 23.2%. The experiment results show that the water-cooling electrome is feasible.
What problem does this paper attempt to address?