Radial Position Sensitive Microchannel Plate Detector with Ring-Strip Resistive Anode and Integrated Electronics

X. Hui,D. Yu,X. Song,W. Wang,J. Liu,X. Li,Y. Zhang,W. Xi,X. Yuan,X. Zhu,B. Ding
DOI: https://doi.org/10.1088/1748-0221/13/09/t09004
2018-01-01
Journal of Instrumentation
Abstract:Using a ring-strip resistive anode and a pair of integrated spectroscopy amplifiers, a circular radius-sensitive microchannel plate (MCP) detector is developed for future studies of ion-atom collisions. The detector is tested by a Fe-55 X-ray source and a radial resolution of 0.2 mm is achieved.
What problem does this paper attempt to address?