Detection Efficiency and Spatial Resolution Analysis of a Nmcp Detector

Qiangwei Liu,Yujie Zhao,Yangyi Yu,Yigang Yang,Zhi Zhang
DOI: https://doi.org/10.1109/nssmic.2017.8532771
2017-01-01
Abstract:Neutron sensitive microchannel plates (nMCP), as key components of image intensifier based neutron energy selective imaging systems, have played major roles in the neutron energy selective imaging and have been applied in many areas for its high detection efficiency and superb spatial resolution. Doping microchannel plate with neutron sensitive nuclides, proposed by G.W. Fraser et al., is a typical method to make normal MCP become neutron sensitive. Inspired by this approach, microchannel plates doped with 3 mol% natGd 2 O 3 have been developed in this study, both in 50 mm and 10 6 mm diameter. Experimental results showed the neutron detection efficiency@25.3 meV was 34%, with the spatial resolution of 64 μm and 88 μm for 15 kV X-rays and 25.3 meV neutrons, respectively. Both the detection efficiency and spatial resolution are intrinsically determined by the internal processes occurring in the nMCP, such as the neutron absorption, IC electron emission and ionization, the electron multiplication and the electron cloud collection, all of which will somehow determine the performance of the nMCP detector. This paper aims to thoroughly describe these processes and investigate their influences on the detection efficiency and spatial resolution of the nMCP detector.
What problem does this paper attempt to address?