Position-sensitive plastic scintillator detector with WLS-fiber readout

Jia-Ning Dong,Yun-Long Zhang,Zhi-Yong Zhang,Dong Liu,Zi-Zong Xu,Xiao-Lian Wang,Shu-Bin Liu
DOI: https://doi.org/10.1007/s41365-018-0449-2
2018-01-01
Nuclear Science and Techniques
Abstract:In this paper, the design, construction, simulation, and performance of a position-sensitive plastic scintillator detector are presented. The readout of the detector uses wavelength-shifting fibers coupled with multi-anode photomultipliers (PMTs) for the x - and y -dimensions. After calibrating the multi-anode PMTs, a two-dimensional projection image of the square scintillator telescope hit by cosmic muons is demonstrated. By performing a cosmic test with the Micromegas telescope, the position resolution of the detector was determined to reach approximately 8.6 mm, which is close to the value obtained in physical simulation.