Design and Optimization of an Energy Degrader with a Multi-Wedge Scheme Based on Geant4

Zhikai Liang,Kaifeng Liu,Bin Qin,Wei Chen,Xu Liu,Dong Li,Yongqian Xiong
DOI: https://doi.org/10.1016/j.nima.2018.01.073
2018-01-01
Abstract:A proton therapy facility based on an isochronous superconducting cyclotron is under construction in Huazhong University of Science and Technology (HUST). To meet the clinical requirements, an energy degrader is essential in the beamline to modulate the fixed beam energy extracted from the cyclotron. Because of the multiple Coulomb scattering in the degrader, the beam emittance and the energy spread will be considerably increased during the energy degradation process. Therefore, a set of collimators is designed to restrict the increase in beam emittance after the energy degradation. The energy spread will be reduced in the following beam line which is not discussed in this paper. In this paper, the design considerations of an energy degrader and collimators are introduced, and the properties of the degrader material, degrader structure and the initial beam parameters are discussed using the Geant4 Monte-Carlo toolkit, with the main purpose of improving the overall performance of the degrader by multiple parameter optimization.
What problem does this paper attempt to address?