Calibration of an Ej309 Liquid Scintillator Using an Ambe Neutron Source

Huaiyong Bai,Zhimin Wang,Luyu Zhang,Yi Lu,Haoyu Jiang,Jinxiang Chen,Guohui Zhang
DOI: https://doi.org/10.1016/j.nima.2017.04.028
2017-01-01
Abstract:As the light output function of the EJ309 liquid scintillator is nonlinear with respect to the energy of the recoil proton and it is difficult to be predicted, it need to be measured. In the present work, an EJ309 liquid scintillator, 5.1cm in diameter and 5.1cm in length, is calibrated with an AmBe neutron source using the time-of-flight method. A Monte Carlo simulation is performed to improve the accuracy of the measured light output function. The obtained light output function is presented and it is compared with existing measurements. The energy resolution of the EJ309 Liquid scintillator is also illustrated.
What problem does this paper attempt to address?