Neutron Emission Measurement At The Hl-2a Tokamak Device With A Liquid Scintillation Detector

Xufei Xie,Zhongjing Chen,Xingyu Peng,Xi Yuan,Xing Zhang,Giuseppe Gorini,Zhiqiang Cui,Tengfei Du,Zhimeng Hu,Tao Li,Tieshuan Fan,Jinxiang Chen,Xiangqing Li,Guohui Zhang,Guoliang Yuan,Jinwei Yang,Qingwei Yang
DOI: https://doi.org/10.1063/1.4896120
IF: 1.6
2014-01-01
Review of Scientific Instruments
Abstract:Neutron emission measurement at the HL-2A tokamak device with a liquid scintillation detector is described. The detector was placed at a location with little structure material in the field of view, and equipped with a gain monitoring system which could provide the possibility to evaluate the gain variation as well as to correct for the detector response. Time trace of the neutron emissivity was obtained and it was consistent with the result of a standard U-235 fission chamber. During the plasma discharge the neutron yield could vary by about four orders of magnitude and the fluctuation of the detector gain was up to about 6%. Pulse height spectrum of the liquid scintillation detector was constructed and corrected with the aid of the gain monitoring system, and the correction was found to be essential for the assessment of the neutron energy spectrum. This successful measurement offered experience and confidence for the application of liquid scintillation detectors in the upcoming neutron camera system. (C) 2014 AIP Publishing LLC.