Reduction of the Polarization Sensitivity of Grating-Based Spectrometer Using an Achromatic Quarter-Wave Plate

Yajun Pang,Yinxin Zhang,Huaidong Yang,Zhanhua Huang
DOI: https://doi.org/10.1016/j.optcom.2017.02.044
IF: 2.4
2017-01-01
Optics Communications
Abstract:Spectrometers are generally sensitive to the polarization state of the measured light, which affects their measurement accuracy. In this study, we developed a method to reduce the polarization sensitivity by using an achromatic quarter-wave plate in a double-pass optical structure. By rotating the polarization state of the second pass beam with an achromatic quarter-wave plate, the diffraction efficiencies of the grating for the s-polarized and p-polarized light become nearly the same. The formulas of this polarization-insensitive system were strictly derived. We designed a double-pass spectrometer with a grating of 1050 lines/mm and a reflective achromatic quarter-wave plate. The results of the numerical simulation show that the polarization-dependent loss of the designed spectrometer is less than 0.5 dB.
What problem does this paper attempt to address?