Time-resolved X-ray diffraction diagnostic development for the National Ignition Facility
K. Werellapatha,N. E. Palmer,M. G. Gorman,J. V. Bernier,N. S. Bhandarkar,D. K. Bradley,D. G. Braun,M. Bruhn,A. Carpenter,P. M. Celliers,F. Coppari,M. Dayton,C. Durand,J. H. Eggert,B. Ferguson,B. Heidl,C. Heinbockel,R. Heredia,J. Huckins,E. Hurd,W. Hsing,C. M. Krauland,A. E. Lazicki,D. Kalantar,J. Kehl,K. Killebrew,N. Masters,M. Millot,S. R. Nagel,R. B. Petre,Y. Ping,D. N. Polsin,S. Singh,C. V. Stan,D. Swift,J. Tabimina,A. Thomas,T. Zobrist,L. R. Benedetti
DOI: https://doi.org/10.1063/5.0161343
IF: 1.6
2024-01-01
Review of Scientific Instruments
Abstract:We present the development of an experimental platform that can collect four frames of x-ray diffraction data along a single line of sight during laser-driven, dynamic-compression experiments at the National Ignition Facility. The platform is comprised of a diagnostic imager built around ultrafast sensors with a 2-ns integration time, a custom target assembly that serves also to shield the imager, and a 10-ns duration, quasi-monochromatic x-ray source produced by laser-generated plasma. We demonstrate the performance with diffraction data for Pb ramp compressed to 150 GPa and illuminated by a Ge x-ray source that produces ∼7 × 1011, 10.25-keV photons/ns at the 400 μm diameter sample.
instruments & instrumentation,physics, applied