Assessment of Regional Myocardial Blood Flow by Self-Made Microbubble Contrast Agent

刘诗珍,舒先红,潘翠珍,姚瑞明,沈学东,陈灏珠
DOI: https://doi.org/10.3760/j.issn:1004-4477.2002.04.012
2002-01-01
Abstract:Objective Using self-made microbubble contrast agent to quantify the change of the regional myocardial blood flow (MBF) with myocardial contrast echocardiography (MCE). Methods ①Sonicated the mixture of 1% albumin, 5% dextrose and perfluorocarbon gases using 70, 140 and 210 seconds, and compared the size and concentration of the three kind of microbubbles. ②MCE was performed in 6 open chest dogs during baseline, hyperemia by dipyridamole and different grades of coronary stenosis and regional MBF was calculated. Results ①The microbubble size of the 3 groups was ( 9.01 ± 3.14 )μm, ( 6.45 ± 2.39 )μm, ( 4.86 ± 2.09 )μm, respectively, and the concentration was ( 1.71 ± 0.22 )×10 8 microbubbles/ml, ( 3.77 ± 0.92 )×10 8 microbubbles/ml, ( 8.95 ± 0.81 )×10 8 microbubbles/ml, respectively. And group 3 was used. ②Myocardial bed could be completely fulfilled by the contrast agent when coronary stenosis was none or mild, although the time needed was different. ③MBF derived from MCE had good correlations with those derived from radiolabeled microspheres ( Y = 0.76 X + 0.07 , r = 0.82 , P = 0.01 ). Conclusions MCE with self-made microbubble contrast agent can quantifiably reflect regional MBF.
What problem does this paper attempt to address?